This Averaging Axial Extensometer (Series 632.17) is designed to simultaneously measure the axial deflection on opposite sides of the specimen. This extensometer is typically used to perform testing on solid or tubular composite or metal specimens.
Simultaneously measures axial deflection on opposite sides of the specimen to reduce measurement errors caused by specimen bending
Securely attaches the extensometer to the specimen while providing the necessary freedom of movement
Ensures true center-point bending throughout the entire travel range of the extensometer
Accommodates specimen bending or twisting without slippage or output distortion, increasing accuracy
The Series 632.17 Averaging Axial extensometer is designed to simultaneously measure the axial deflection on opposite sides of the specimen. The extensometer is typically used to perform testing on solid or tubular composite or metal specimens.
The extensometer consists of two sensor units, one on either side of the specimen. The sensor units measure the axial deflection over the gage length of the specimen. Two contact points, connected to each sensor unit assembly, contact the specimen. The contact points are made of hardened steel or (as an option) carbide. Both materials are available with conical point contact ends or vee-chisel contact ends.
Mounting
Signal Output
Accuracy
The 632.17 Averaging Axial extensometer is designed for extremely accurate strain measurements. Exceptional accuracy and linearity are provided by the cross flexure design, which ensures true center-point bending throughout the entire travel range of the extensometer.
Conditioning Electronics
The extensometer is used with MTS conditioning electronics or other conditioning electronics capable of providing an excitation voltage of 8 volts (not to exceed 12 volts). The extensometer output is approximately 3.4 millivolts per volt of excitation. Bridge resistance is 500 ohms for option 001 (averaged output) and 1000 ohms for option 002 (individual output).
| Metric Units | |||||||
| MODEL NUMBER | GAGE LENGTH | MEASURING RANGE (% STRAIN) | HEIGHT | LENGTH | DEPTH | TEMPERATURE RANGE MIN/MAX* | PART NUMBER |
| 632.17F-x0 | |||||||
| 632.17F-40 | 10 ±0.05 mm | +12% / -5% | 63.5 mm | 101.6 mm | 38.1 mm | -100°C / 150°C | 047-404-802 |
| 632.17F-20 | 25 ±0.05 mm | +5% / -2% | 63.5 mm | 101.6 mm | 38.1 mm | -100°C / 150°C | 047-401-102 |
| US Customary Units | |||||||
| MODEL NUMBER | GAGE LENGTH | MEASURING RANGE (% STRAIN) | HEIGHT | LENGTH | DEPTH | TEMPERATURE RANGE MIN/MAX* | PART NUMBER |
| 632.17E-x0 | |||||||
| 632.17E-30 | 0.5 ±0.002 in | +10% / -4% | 2.5 in | 4.0 in | 1.5 in | -150°F / 300°F | 047-404-801 |
| 632.17E-20 | 1.0±0.002 in | +5% / -2% | 2.5 in | 4.0 in | 1.5 in | -150°F / 300° F | 047-401-101 |
| *These units may be used at 25°C (50 °F) higher than the listed temperatures, but only for durations less than 24 hours. | |||||||