Even simpler to use than a traditional contacting extensometer, the MTS Advantage™ Optical Extensometer (AOX) also provides all the precision and productivity advantages of a non-contacting strain measurement solution. Whether you are measuring QC tensile strength, long-term fatigue or crack growth, this extensometer can accommodate everything from fine resolution measurements to high-elongation requirements and more.
Real-time data rate of 300 - 3000 Hz
No grids or bars required for pre-test calibration
Instant Reset feature improves productivity
Once set up, it's always measuring
The MTS Advantage Optical Extensometer performance combines the latest in optical and algorithmic measurement techniques to deliver the high resolution, data rates and accuracy needed for uniaxial testing.
Resolution
<0.5 μm quasistatic
<2.5 μm dynamic
<0.1 μm creep; typical RMS resolution at typical settings. Resolution is a function of the optics and filter settings.
Extensometer Accuracy Class
ISO 9513 Class 0,5 and ASTM E83 Class B-1 or better, typical, for gage lengths ≥10 mm
Real-time data rate: 300-3000 Hz
As a result, the MTS Advantage Optical Extensometer is equipped with the accuracy and fine resolution required to measure modulus, offset yield, stress-strain curves, and strain at failure for both low and high modulus materials.
The MTS AOX models with telecentric lenses are engineered to provide the accuracy and fine resolution required for testing high-modulus materials such as metals, composites, ceramics, and CMCs:
The MTS AOX-500CE optics option can measure more than 1000% strain (500 mm field-of-view and 10 or 25 mm gage length):
Non-contact strain measurement is the best way, and sometimes the only way, to measure strain with delicate, fine, and thin specimens. Use the AOX extensometer to measure modulus, offset yield, stress-strain curves, and elongation to break with wire, metal foil, polymer film, high-modulus fibers, and polymer filaments:
Specifications |
|
Field-of-view | 50-500+ mm FoV Optics Packages; other ranges available – contact MTS |
Real-Time Data Rate | 300-3000 Hz; includes image frame rate, analog output, and digital output |
Strain Outputs: Analog Output | ±10V, short-protected, selectable units and range; includes 2.4 m (8 ft) shielded output cable |
Digital Output | 16-64 bit over RS232 with selectable units & data type; 2900 strain readings per second Typical data rate @32 bits; includes 1.8 m (6 ft) shielded null modem cable
|
Resolution | <0.5 μm quasistatic <2.5 μm dynamic <0.1 μm creep; typical RMS resolution at typical settings. Resolution is a function of the optics and filter settings. |
Extensometer Accuracy Class | ISO 9513 Class 0.5 and ASTM E83 Class B-1 or better typical at ≥10 mm gage length |
Absolute Accuracy | PT models: <±1 μm or 0.5% of value, not exceeding ±30 μm typical CE models: <±1 μm or 0.5% of value |
Minimum Specimen Size | < Ø20 µm (0.001”) |
Gage Length | 2 mm minimum, ≥4x specimen width or diameter recommended; for smaller sizes contact MTS |
Maximum Elongation | >1000%, limited by field-of-view and gage length |
Maximum Tracking Speed | >1500 mm/second (90,000 mm/minute) |
Cyclic Testing | >100 Hz cyclic test frequency, waveform independent |
Strain Control | Suitable for monotonic and cyclic strain control applications |
Out-of-Plane Sensitivity | Allowable out-of-plane motion for ISO 9513 Class 0,5 @15 mm gage length and ASTM E83 Class B-1: <1000 µm (0.040”) for AOX-PT-xx <25 µm (0.001”) for AOX-CE-xx |
Power Supply | 100-240 VAC, 50-60 Hz, 100W, IEC 320 C14 receptacle. Specify plug type when ordering |
System Environment (environments) | 10-40°C (50-100°F), for use and storage; 20-80% relative humidity non-condensing |
Host PC (optional): | Requires Windows 10 PC, 900×550 minimum display, one serial or USB port |
|
|
Options |
|
|
|
Specimen marking is simple and fast. Templates and paint pens are included with every Optics Package.